STATus Subsystem
STATus:QUEStionable:CALibration[:SUMMary]
HP E4418B/E4419B Programming Guide 10-21
STATus:QUEStionable:CALibration[:SUMMary]
The questionable calibration summary register set contains bits which
give an indication of the quality of the data produced by the power meter
due to its calibration status.
The following bits in these registers are used by the power meter:
These bits are set by the following:
Error -231, “Data questionable; CH<A|B>:ZERO ERROR”
Error -231, “Data questionable; CAL ERROR”
Error -231, “Data questionable; CAL ERROR ChA
Error -231, “Data questionable; CAL ERROR ChB”
These bits are cleared when any of the three commands listed above
succeed and no errors are placed on the error queue.
Syntax
Bit
Number
Decimal
Weight
Definition
0 - Not used
1 2 Summary of Channel A CALibration
2 4 Summary of Channel B CALibration (HP E4419B only)
3-15 - Not used
STAT
:QUES
:CAL
:SUMM