prog.book : ch9sens.fm 5 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
HP E4418B/E4419B Programming Guide 9-5
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
This command is used to enter a value for the filter length. If
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO is set to ON then entering a
value for the filter length automatically sets it to OFF. Increasing the value
of filter length increases measurement accuracy but also increases the
time taken to make a power measurement.
Entering a value using this command automatically turns the
[SENSe[1]]|SENSe2:AVERage:STATe command to ON.
Note For most applications, automatic filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt:AUTO ON) is the best
mode of operation. However, manual filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>) is
useful in applications requiring either high resolution or fast
settling times.
Syntax
:COUN
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
SENS
1
AVER
SENS2
: