prog.book : ch9sens.fm 52 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <Boolean>
9-52 HP E4418B/E4419B Programming Guide
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <Boolean>
This command is used with the HP E-series power sensors to enable and
disable autoranging. When autoranging is ON, the power meter selects the
best measuring range for the measurement. When autoranging is set to
OFF, the power meter remains in the currently set range.
The [SENSe[1]]|SENSe2:POWer:AC:RANGe command disables
autoranging.
If INITiate:CONTinuous is set to ON and TRIGger:SOURce is set to
IMMediate, the range will track the input power if
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO is ON.
If the power meter is not making measurements then autoranging only
occurs when the power meter is triggered.
Syntax
Example
Reset Condition
On reset, autoranging is enabled.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO?
The query enters a 1 or 0 into the output buffer indicating the status of
autoranging.
• 1 is returned when autoranging is enabled.
• 0 is returned when autoranging is disabled.
POW:AC:RANG:AUTO 0 This command disables
autoranging.
SENS
1
:AC
POW
SENS2
:
:AUTO
0|OFF
1|ON
Space
?
:RANG

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals