prog.book : ch7mem.fm 10 Mon Feb 15 09:24:37 1999
MEMory Subsystem
MEMory:CLEar[:NAME] <string>
7-10 HP E4418B/E4419B Programming Guide
MEMory:CLEar[:NAME] <string>
This command is used to clear the contents of a given sensor calibration
table, frequency dependent offset table, or save/recall register.
Although the table remains, a
MEMory:TABLe:FREQuency|GAIN:POINts? query returns a 0 as there
are no contents in the table.
For sensor calibration tables and frequency dependent offset tables, this
command is an alternative form of the MEMory:CLEar:TABLE command,
the only difference being the method in which the table is selected.
Note The contents cleared using this command are non-recoverable.
Syntax
Parameters
Example
Error Messages
If the table or save/recall register name does not exist, error -224, “Illegal
parameter value” occurs.
Item Description/Default Range of Values
string
String data
.
Any existing table
name or
save/recall register.
MEM:CLE "HP8485A" This command clears the contents
of sensor calibration table
HP 8485A
MEM
:CLE
:NAME
Space
string

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals