prog.book : ch9sens.fm 23 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <Boolean>
HP E4418B/E4419B Programming Guide 9-23
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <Boolean>
This command is used to enable and disable the pulse power
measurement feature.
The [SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 command
should be used to enter the duty cycle of the signal you want to measure.
Syntax
Example
Reset Condition
On reset, the pulse power measurement feature is disabled.
Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe?
The query enters a 1 or 0 into the output buffer indicating the status of
the pulse power measurement feature.
• 1 is returned when the pulse power measurement feature is
enabled.
• 0 is returned when the pulse power measurement feature is
disabled.
CORR:DCYC:STAT 1 This command enables the pulse
measurement feature on
channel A.
SENS
1
:DCYC
CORR
SENS2
:
:STAT
0|OFF
1|ON
Space
?
:GAIN3