prog.book : ch9sens.fm 24 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <Boolean>
9-24 HP E4418B/E4419B Programming Guide
Query Example
Error Messages
If [SENSe[1]]|SENSe2:CORRection:DCYCle:STATus is set to ON while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, “Settings
Conflict” occurs.
If this command is used when an HP E-series power sensor is connected,
the error -310, “System error;Dty Cyc may impair accuracy with ECP
sensor” occurs. If you are using the HP E4419B the error message
specifies the channel.
CORR:GAIN3:STAT? This command queries whether
the pulse measurement feature is
on or off.