EasyManua.ls Logo

HP E4418B - Page 307

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
prog.book : ch9sens.fm 6 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
9-6 HP E4418B/E4419B Programming Guide
Parameters
Example
Reset Condition
On reset, the filter length is set to 4.
Query
AVERage:COUNt? [MIN|MAX]
The query returns the current setting of the filter length or the values
associated with MIN and MAX. The format of the response is <NR1>.
Query Example
Error Message
If a filter length value is entered using
[SENSe[1]]|SENSe2:AVERage:COUNt while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, “Settings
Conflict” occurs. However, the filter length value is set but the
[SENSe[1]]|SENSe2:AVERage:STATe command is not automatically
set ON.
Item Description/Default Range of Values
numeric_value A numeric value defining the filter length.
The default value, DEF, is 4. The value for
MIN is 1. The value for MAX is 1024.
1 to 1024
DEF
MIN
MAX
AVER:COUN 400 This command enters a filter
length of 400 for channel A.
AVER:COUN? This command queries the filter
length for channel A.

Table of Contents

Other manuals for HP E4418B

Related product manuals