prog.book : ch9sens.fm 6 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
9-6 HP E4418B/E4419B Programming Guide
Parameters
Example
Reset Condition
On reset, the filter length is set to 4.
Query
AVERage:COUNt? [MIN|MAX]
The query returns the current setting of the filter length or the values
associated with MIN and MAX. The format of the response is <NR1>.
Query Example
Error Message
If a filter length value is entered using
[SENSe[1]]|SENSe2:AVERage:COUNt while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, “Settings
Conflict” occurs. However, the filter length value is set but the
[SENSe[1]]|SENSe2:AVERage:STATe command is not automatically
set ON.
Item Description/Default Range of Values
numeric_value A numeric value defining the filter length.
The default value, DEF, is 4. The value for
MIN is 1. The value for MAX is 1024.
1 to 1024
DEF
MIN
MAX
AVER:COUN 400 This command enters a filter
length of 400 for channel A.
AVER:COUN? This command queries the filter
length for channel A.

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals