prog.book : ch4cal.fm 13 Mon Feb 15 09:24:37 1999
CALibration Subsystem
CALibration[1|2]:RCFactor <numeric_value>
HP E4418B/E4419B Programming Guide 4-13
Query
CALibration[1|2]:RCFactor? [MIN|MAX]
The query returns the current setting of the reference calibration factor or
the values associated with MIN and MAX.
Query Example
Error Message
If this command is used when an HP E-series power sensor is connected
the error -241, “Hardware missing” occurs.
CAL2:RCF? This command queries the
reference calibration factor of
channel B.