EasyManua.ls Logo

HP E4418B - Page 290

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
prog.book : ch7mem.fm 27 Mon Feb 15 09:24:37 1999
MEMory Subsystem
MEMory:TABLe:GAIN[:MAGNitude] <numeric_value>{,<numeric_value>}
HP E4418B/E4419B Programming Guide 7-27
Parameters
Example
Query
MEMory:TABLe:GAIN[:MAGNitude]?
The query returns a list of calibration factor/offset points for the currently
selected table.
Query Example
Error Messages
If more than 81 calibration factors for sensor calibration tables, or 80
offsets for frequency dependent offset tables are in the list, error -108,
“Parameter not allowed” occurs.
If a table is not specified using the MEMory:TABLe:SELect command, the
data cannot be entered and error -221, “Settings conflict” occurs.
If any of the calibration/offset factors are outwith the allowed range, error
-222, “Data out of range” occurs.
Item Description/Default Range of Values
numeric_value A numeric value for the
calibration/offset factors. The units are
PCT.
1.0 to 150.0
MEM:TABL:SEL "Sensor_1"
MEM:TABL:GAIN 97,99.5,97.4
This command enters a reference
calibration factor of 97% and
calibration factors of 99.5% and
97.4% into the sensor calibration
table.
MEM:TABL:GAIN? This command queries the
calibration factor/offset in the
current table.

Table of Contents

Other manuals for HP E4418B

Related product manuals