prog.book : ch9sens.fm 8 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <Boolean>|ONCE
9-8 HP E4418B/E4419B Programming Guide
Note For most applications, automatic filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt:AUTO ON) is the best
mode of operation. However, manual filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>) is
useful in applications requiring either high resolution or fast
settling times.
Syntax
Example
Reset Condition
On reset, automatic averaging is enabled.
Query
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO?
The query enters a 1 or 0 into the output buffer indicating whether
automatic filter length is enabled or disabled.
1 is returned when automatic filter length is enabled.
0 is returned when automatic filter length is disabled.
Query Example
AVER:COUN:AUTO OFF This command disables
automatic filter length selection for
channel A.
AVER:COUN:AUTO? This command queries whether
automatic filter length selection is
on or off for channel A.
SENS
1
:COUN
AVER
0|OFF
1|ON
Space
?
SENS2
:AUTO
:
ONCE

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals