prog.book : ch4cal.fm 8 Mon Feb 15 09:24:37 1999
CALibration Subsystem
CALibration[1|2]:AUTO <Boolean>|ONCE
4-8 HP E4418B/E4419B Programming Guide
Query
CALibration[1|2]:AUTO?
The query always returns a value of 0.
Error Message
If the calibration was not carried out successfully the error -231, “Data
Questionable; CAL ERROR” occurs. If you are using the HP E4419B the
error message specifies on which channel the calibration failed.
If this command is set to ON the error -224, “Illegal parameter value”
occurs.