Page #32 background image
prog.book : chapter1.fm 8 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Configuring the Remote Interface
1-8 HP E4418B/E4419B Programming Guide
1. This command is accepted but has no active function.
2. This command is not an original HP 437B command. However, it can be used to
set the measurement speed to 20 or 40 readings/sec in HP 437B mode. See
SENSE:SPEED for more details.
3. This command is not an original HP 437B command. However, it can be used to
terminate the HP 437B language and select the SCPI language. Note that it is
recommended that the instrument is Preset following a language switch.
4. This command is not an original HP 437B command. However, it can be used to
allow the last measurement result to be transmitted. This is equivalent to
sending the power meter talk address in HP-IB mode to fetch the last reading
(provided no query is pending).
5. Always returns 0000 in HP 437B language.
SP
1
Special
SPD 20|40
2
20 or 40 readings/sec
*SRE
Set the service request mask
*SRE?
Service request mask query
ST
Store (save) power meter configuration
*STB?
Read status byte
SYST:LANG SCPI
3
Selects SCPI language
TK?
4
Last measurement result transmitted
TR0
Trigger hold
TR1
Trigger immediate
TR2
Trigger with delay
TR3
Trigger free run
*TST?
5
Self test query
UP
1
Up arrow key
ZE
Zero
@1
Prefix for status mask
@2
Learn mode prefix
%
Percent
Command Description

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals