prog.book : ch9sens.fm 21 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut][:MAGNitude]
HP E4418B/E4419B Programming Guide 9-21
Syntax
Parameters
Example
Reset Condition
On reset, the duty cycle is set to 1% (DEF).
Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut][:MAGNitude]? [MIN|MAX]
The query returns the current setting of the duty cycle or the values
associated with MIN and MAX.
Item Description/Default Range of Values
numeric_value A numeric value for the duty cycle. The
default value, DEF, is 1%. The value for
MIN is 0.001%. The value for MAX is
99.999%. The units are PCT, and are
optional.
0.001 to 99.999PCT
DEF
MIN
MAX
CORR:DCYC 90PCT This command sets a duty cycle of
90% for channel A.
SENS
1
:DCYC
CORR
SENS2
:
:INP
:MAGN
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:GAIN3