SENSe Subsystem
[SENSe[1]]|SENSe2:LIMit:CLEar:AUTO <Boolean>|ONCE
HP E4418B/E4419B Programming Guide 9-41
Query
[SENSe[1]]|SENSe2:LIMit:CLEar:AUTO?
The query command enters a 1 or 0 into the output buffer indicating
whether limit failures are cleared automatically when a new
measurement is initiated on the specified channel.
• 1 is entered into the output buffer when limit failures are cleared
automatically when a new measurement is initiated.
• 0 is entered into the output buffer when limit failures are not
cleared automatically when a new measurement is initiated.
In the case where limit failures are cleared once, when a query occurs a 1
is entered into the output buffer if no measurement is initiated. If a
measurement is initiated then 0 is entered.
Query Example
SENS1:LIM:CLE:AUTO? This command queries when the
FCO is cleared for channel A.