SENSe Subsystem
[SENSe[1]]|SENSe2:SPEed <numeric_value>
HP E4418B/E4419B Programming Guide 9-55
Parameters
Example
Reset Condition
On reset, the speed is set to 20 readings/second.
Query
[SENSe[1]]|SENSe2:SPEed?
The query returns the current speed setting, either 20, 40 or 200.
Query Example
Error Message
If <speed> is not 20, 40 or 200, error -224, “Illegal parameter value”
occurs.
If an HP E-series power sensor is not connected and the
<numeric_value> is set to 200 readings/second the error -241,
“Hardware missing” occurs.
Item Description/Default Range of Values
numeric_value A numeric value for the measurement
speed. The default is 20.
20
40
200
SPE 40 This command sets the channel A
speed to 40 readings/second.
SPE? This command queries the current
speed setting for channel A.