prog.book : ch11syst.fm 23 Mon Feb 15 09:24:37 1999
SYSTem Subsystem
SYSTem:COMMunicate:SERial:TRANsmit:PARity[:TYPE] EVEN|ODD|ZERO|ONE|NONE
HP E4418B/E4419B Programming Guide 11-23
Syntax
Example
Reset Condition
On reset, the parity selection is unaffected.
Query
SYSTem:COMMunicate:SERial:TRANsmit:PARity[:TYPE]?
The query returns the current setting of the transmit/receive parity type
(EVEN, ODD, ZERO, ONE or NONE).
Query Example
SYST:COMM:SER:TRAN:PAR ODD This command sets the transmitter
to include an odd parity bit in the
transmitted data and the receiver
to check for odd parity in the
received data.
SYST:COMM:SER:TRAN:PAR? This command queries what type
of parity has been set on the
transmitted/received data.
SYST
:COMM
:SER
:PAR
Space
?
EVEN
ODD
ZERO
:TYPE
:TRAN
ONE
NONE