prog.book : ch12trig.fm 2 Mon Feb 15 09:24:37 1999
TRIGger Subsystem
TRIGger Subsystem
12-2 HP E4418B/E4419B Programming Guide
TRIGger Subsystem
The TRIGger subsystem is used to synchronize device actions with events.
The TRIGger subsystem includes the ABORt, INITiate and TRIGger
commands. These are all at the root level in the command hierarchy but
they are grouped here because of their close functional relationship.
Further information on triggering is detailed in “Triggering the Power
Meter”, on page 1-61.
The numeric suffix of the TRIGger program mnemonic in the TRIGger
commands refers to a channel, that is TRIGger1 and TRIGger2 represent
channel A and channel B respectively. If you are using the single channel
HP E4418B power meter the TRIGger2 commands are irrelevant and
cause the error “Header suffix out of range.
Keyword Parameter Form Notes Page
ABORt[1|2] [no query] page 12-3
INITiate[1|2]
:CONTinuous <Boolean> page 12-5
[:IMMediate] [no query] page 12-5
TRIGger[1|2]
:DELay
:AUTO <Boolean> page 12-9
[:IMMediate] [no query] page 12-11
:SOURce BUS|IMMediate|HOLD page 12-12

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals