prog.book : chapter1.fm 24 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Making Measurements
1-24 HP E4418B/E4419B Programming Guide
Using INITiate and FETCh?
Some fine tuning of measurements can be carried out using the
CONFigure command and INITiate and FETCh? commands. For
example, in the above program segment some fine tuning can be carried
out by setting the filter length to 1024 and the trigger delay off.
ABOR2
CONF1 -50,DEF,(@2)
SENS2:AVER:COUN 1024
TRIG2:DEL:AUTO OFF
INIT2
FETC1?
ABOR2 Aborts channel B
CONF1 -50,DEF,(@2) Configures the upperwindow to
make a channel B measurement
using an expected power of
-50 dBm and the current
resolution setting.
INIT2 Causes channel B to make a
measurement.
FETC1? Retrieves the upper window’s
measurement.