prog.book : chapter1.fm 26 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Making Measurements
1-26 HP E4418B/E4419B Programming Guide
The following program segment assumes that channel A is currently being
measured on the upper window.
Some fine tuning of the above program segment can be carried out for
example, by setting the trigger delay off.
ABOR1
CONF1 DEF,3
TRIG1:DEL:AUTO OFF
INIT1:IMM
FETC1?
ABOR1 Aborts channel A.
CONF1 DEF,3 Configures the upper window to
make a measurement using the
current setting of the expected
power and source list and a
resolution setting of 3.
INIT1 Causes channel A to make a
measurement.
FETC1? Retrieves the upper window’s
measurement.