EasyManuals Logo

HP E4418B Programming Guide

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
prog.book : chapter1.fm 28 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Making Measurements
1-28 HP E4418B/E4419B Programming Guide
Example 6 - Making a Ratio Measurement
The following program segment can be carried out on the HP E4419B. It
queries the lower window to make a ratio measurement of channel A/B.
The expected power level and resolution parameters are defaulted,
leaving them at their current settings. Some fine tuning of the
measurement is carried out by setting the averaging.
Using READ?
ABOR1
ABOR2
CONF2:POW:AC:RAT DEF,DEF,(@1),(@2)
SENS1:AVER:COUN 512
SENS2:AVER:COUN 256
READ2:POW:AC:RAT?
READ2:POW:AC:RAT? DEF,DEF,(@2),(@1) (A second
READ?
query is
sent to make a channel B - channel A ratio measurement using fresh
measurement data.)
Using INITiate and FETCh?
ABOR1
ABOR2
CONF2:POW:AC:RAT DEF,DEF,(@1),(@2)
SENS1:AVER:COUN 512
SENS2:AVER:COUN 256
INIT1:IMM
INIT2:IMM
FETC2:POW:AC:RAT?
FETC2:POW:AC:RAT? DEF,DEF,(@2),(@1) (A second FETCh? query is
sent to make a channel B - channel A measurement using the current
measurement data.)

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals