166 Keysight N9915-90020 User’s Guide
SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)FUTURE
SA Mode Settings
Figure 9-4 FieldFox Antenna Editors – Settings and Values
How to use the Antenna/Cable Editor
1. Press Scale/Amptd > More > Corrections then choose:
— Antenna Off to open the antenna editing softkeys
—
Cable Off to open the cable editing softkeys
2. Then
Edit Antenna or Edit Cable to open the Antenna / Cable Editor.
3. Press
New then Yes to clear all data from the existing Antenna or Cable
table and reset header information to default settings.
4. Then use the arrows to select a field.
— To edit header information, press
Edit then modify the selected field
using the FieldFox labeler.
— Description: Enter a description for the antenna or cable
correction table. This is NOT the filename used to save the
file.
—
Antenna Factor Unit: Determines whether Field Strength
measurement Amplitude units will be available or not, and
how to convert between them. Select None (default
Before editing the antenna settings, it is important to know whether
you are using a magnetic H-field (dB S/m) or electric E-field (dB /m)
antenna. When you setup the Antenna Factor Correction table, you
will select one of these two units so that the FieldFox knows how to
convert the measurement data to various Field Strength Amplitude
units.