Keysight N9915-90020 User’s Guide 767
System SettingsFUTURE
Preset
Continuous
—Hold
is displayed while a single sweep occurs.
How to perform a single sweep while in Hold
1. Press Sweep 3.
2. Then choose one of the following:
Single Automatically sets Continuous OFF (Hold) and causes
FieldFox to make ONE measurement sweep, then hold for the next
Single key press.
—SA Mode ONLY – With
Continuous OFF each successive Run/Hold
press also causes a single sweep.
Continuous ON OFF Makes continuous sweeps. This is the typical
setting when battery power is not critical.
Preset
The Preset key restores the FieldFox measurement settings to a known
condition. You can preset the entire instrument, or just the current Mode (CAT,
SA, and so forth).
When Preset, the FieldFox CalReady calibration is recalled for use in CAT and
NA Modes. Learn more in
“CalReady” on page 127.
In addition to all Preference settings, the following settings always survive
Preset and Mode Preset:
External Reference state (see “Frequency Reference Source” on page 800)
SA Scale and Units (see “Scale and Units” on page 159)
SA Correction states (see
“Field Strength Measurements” on page 163)
VVM Display Resolution (see
“Display Resolution” on page 612)
Marker Peak Excursion and Threshold (see “What Is a ‘Peak’” on page 745)
How to Preset the FieldFox
1. Press Preset.
2. Then choose one of the following:
Preset Returns ALL Modes to the factory default settings.