254 Keysight N9915-90020 User’s Guide
SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)FUTURE
Effective Isotropic Radiated Power (EIRP) – (SA Mode Only)
— Save Horizontal – Saves the current horizontal EIRP value (dBm).
—
Save Vertical – Saves the current vertical EIRP value (dBm).
—
Clear H & V –Clears the saved horizontal and vertical EIRP values
(dBm).
2. Configure Center frequency, Span, Preamp, and RF Attenuation
— Press Freq/Dist
> Center > 2 GHz
— Press
Freq Span > 3 MHz (default)
— Press Scale Amptd
> Preamp > ON (default = OFF)
— Press
RF Atten Auto (default = Auto)
3. Summary of EIRP setup, per step
1.
— Press Rx Ant Gain For this example: 30 dB > Enter.
— Press
Dist to Ant For this example: 100m (default)
— Press
Rx Cable Loss For this example: 2.5 dB
— Press
Margin Comp
4. Start EIRP: Press
Restart
5. When done press:
Save Horizontal and
Save Vertical
6. Optional: To clear the saved horizontal and vertical EIRP values press:
Clear H & V