Keysight N9915-90020 User’s Guide 603
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
Then enter a value using the numeric keypad, the arrows, or the rotary
knob.
Auto Trigger
Used ONLY when Trigger Type = Internal, this setting determines whether the
trigger level is set manually or is set to the default level in the USB Power
Sensor firmware.
—Press Sweep 3
—Then Trigger
—Then Auto Trigger
Then choose from the following:
ON Trigger level is determined by the USB Power Sensor firmware.
OFF The trigger level is set manually using the Trigger Level setting.
Trigger Level
Used when Trigger Type = Internal AND Auto Trigger = OFF (Manual).
—Press Sweep 3
—Then Trigger
—Then
Trig Level
Then set the trigger level using the numeric keypad, the arrows, or the
rotary knob. A line appears on Tr 1 at the current trigger level setting. Data
acquisition occurs when the RF input signal crosses the line. This setting
can be used with either Positive or Negative edge triggering.
Qualification
Use the qualification setting to avoid triggering on false pulses and to aid in
stabilizing the power trace.
The qualification time defines the time the signal has to remain below the
trigger threshold before a transition through the threshold will qualify as a
trigger. This can help stabilize a power trace, particularly with digitally
modulated pulses that may have many transitions through the trigger
threshold. By setting the qualification time longer than the short transitions of
the data signal the sensor will reliably trigger on the pulse and not on the data.
It will also eliminate triggering on very short noise spikes when the trigger level
has to be set to low power.
—Press Sweep 3
It may be may be necessary to set the qualification time to the minimum
20 ns in order to trigger on very short pulses.