Keysight E4980A/AL Precision LCR Meter               121
Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
• During the measurement, the ABORT softkey is shown. Use this key when 
you want to abort open correction 
Step 10. Repeat Step 3 to Step 9 until you have specified all of your desired 
measurement points. 
Step 11. Move the cursor into the OPEN field.
Step 12. Press the ON key to enable the open correction feature. 
To specify frequency points and carry out short correction: 
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Using the cursor keys, select the SPOT No. field.
Step 4. Enter the measurement point number using the softkeys or entry keys. If you 
have used the entry keys to enter the value, the softkey labels change to unit 
labels (x1). 
You can specify up to 201 measurement points. 
Step 5. Using the cursor keys, select the FREQ field.
Step 6. Enter the frequency using the entry keys. When you enter the frequency, the 
softkey labels change to unit labels (Hz, kHz, MHz). 
Step 7. Press the ON softkey.
Step 8. Connect the UNKNOWN terminal and the test fixture and short-circuit the 
HIGH and LOW test terminals.
Step 9. Press the MEAS SHORT softkey. 
• During the measurement, a “SHORT measurement in progress” message is 
shown on the display. 
• When the measurement has finished, the “SHORT measurement in 
progress” message disappears. 
• During the measurement, the ABORT softkey is shown. Use this key when 
you want to abort short correction
Softkey Description
INCR ++ Increments the measurement point number in steps of 10. 
INCR + Increments the measurement point number in steps of 1. 
DECR - Decrements the measurement point number in steps of 1. 
DECR -- Decrements the measurement point number in steps of 10.