EasyManua.ls Logo

Keysight E4980A - Page 124

Keysight E4980A
519 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
124 Keysight E4980A/AL Precision LCR Meter
Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
4-
1. This area is used to:
Carry out correction calculation using either the open/short correction
data generated by the interpolation method or the open/short/load
correction data at specified frequency points.
Which data is used depends on the test frequencies.
Obtain the open/short correction data.
2. This area is used to:
Obtain the open/short/load correction data at each frequency point
specified in the FREQ field.
Enable or disable the open/short/load correction data at each frequency
point specified in the FREQ field.
The following tables show how correction data elements are related to each
other when the correction features are enabled in the OPEN, SHORT, and LOAD
fields. Correction data elements are selected based on the number of
measurement points.
When test frequency point is equal to the frequency point specified in FREQ
field:
When test frequency point is NOT equal to the frequency point specified in
FREQ field:
Correction Frequency point specified in FREQ field
ON OFF
OPEN: ON
SHORT: ON
LOAD: ON
User-specified
User-specified
User-specified
Interpolated
Interpolated
X
Correction Frequency point specified in FREQ field
ON OFF
OPEN: ON
SHORT: ON
LOAD: ON
Interpolated
Interpolated
X
Interpolated
Interpolated
X
User-specifie
d
Correction data at your specified frequency point are used.
Interpolated Correction data generated by interpolation are used.

Table of Contents

Other manuals for Keysight E4980A

Related product manuals