Keysight E4980A/AL Precision LCR Meter 125
Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
X Correction is not performed even when correction features are
enabled in OPEN, SHORT, and LOAD fields.
When you specify a frequency point whose frequency value is identical to
that of an existing frequency point and turn on the correction feature for
both points, the frequency point with the smaller measurement point
number (SPOT No. field) is selected as the correction data.