142                 Keysight E4980A/AL Precision LCR Meter 
Configuring Measurement Conditions (Display and Function Related Settings)
LIMIT TABLE SETUP Page
4-
Step 6. Move the cursor into the 2nd LOW field and enter the lower limit value for the 
secondary parameter. 
Step 7. Move the cursor into the 2nd HIGH field and enter the upper limit value for 
the secondary parameter. 
To configure the limit values for sequential mode: 
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Using the cursor keys, select the BIN 1 LOW field.
Step 4. Enter the limit value using the entry keys. When you enter the value, the 
softkey labels change to unit labels (p, n, u, m, x1). 
If you want to clear your selected limit value, press the CLEAR softkey; also, 
you can clear all the limit values for your selected bin by pressing the CLEAR 
LINE softkey. 
Step 5. Repeat Step 4 using the cursor keys until you have entered the upper limit for 
bin 9 (BIN 9 HIGH). 
Step 6. Move the cursor into the 2nd LOW field and enter the Lower limit value for 
the secondary parameter. 
Step 7. Move the cursor into the 2nd HIGH field and enter the upper limit value for 
the secondary parameter. 
To clear the entire table: 
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Using the cursor keys, select the BIN field.
Step 4. Press the CLEAR TABLE softkey.
The LOW field accepts an entry for bin1 only and does not accept an entry 
for bins 2 through 9. The value entered into the HIGH field for bin 1 is used 
as the lower limit for bin 2, but the LOW field for bin 2 displays no value.
You can enter the lower or upper limit values using the HIGHx(-1) or 
LOWx(-1) softkey.