Remote Commands
R&S
®
FSW
1028User Manual 1173.9411.02 ─ 43
Manual operation: See "Zero Span" on page 133
See "Span" on page 444
See "Zero Span" on page 445
See "Last Span" on page 445
[SENSe:]FREQuency:SPAN:FULL
This command restores the full span.
Manual operation: See "Full Span" on page 445
[SENSe:]FREQuency:STARt <Frequency>
This command defines a start frequency for measurements in the frequency domain.
Parameters:
<Frequency> 0 to (fmax - min span)
*RST: 0
Default unit: HZ
Example:
FREQ:STAR 20MHz
Manual operation: See "Frequency Sweep" on page 133
See "Start / Stop" on page 445
[SENSe:]FREQuency:STOP <Frequency>
This command defines a stop frequency for measurements in the frequency domain.
Parameters:
<Frequency> min span to fmax
*RST: fmax
Default unit: HZ
Example:
FREQ:STOP 2000 MHz
Manual operation: See "Frequency Sweep" on page 133
See "Start / Stop" on page 445
14.7.1.2 Configuring Signal Tracking
When signal tracking is activated, the maximum signal is determined after each fre-
quency sweep and the center frequency is set to the frequency of this signal. Thus with
drifting signals the center frequency follows the signal.
For more details see Chapter 8.3.1, "Impact of the Frequency and Span Settings",
on page 441..
CALCulate<n>:MARKer<m>:FUNCtion:STRack[:STATe]..................................................1029
CALCulate<n>:MARKer<m>:FUNCtion:STRack:BANDwidth............................................ 1029
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold.............................................1029
CALCulate<n>:MARKer<m>:FUNCtion:STRack:TRACe.................................................. 1030
Setting Basic Measurement Parameters

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals