Remote Commands
R&S
®
FSW
1081User Manual 1173.9411.02 ─ 43
[SENSe:]MIXer<x>:LOSS:HIGH <Average>
This command defines the average conversion loss to be used for the entire high (sec-
ond) range.
Suffix:
<x>
.
irrelevant
Parameters:
<Average> numeric value
Range: 0 to 100
*RST: 24.0 dB
Default unit: dB
Example:
MIX:LOSS:HIGH 20dB
Manual operation: See "Conversion Loss" on page 422
[SENSe:]MIXer<x>:LOSS:TABLe:HIGH
<FileName>
This command defines the file name of the conversion loss table to be used for the
high (second) range.
Suffix:
<x>
.
irrelevant
Parameters:
<FileName> String containing the path and name of the file.
Example:
MIX:LOSS:TABL:HIGH 'MyCVLTable'
Manual operation: See "Conversion Loss" on page 422
[SENSe:]MIXer<x>:LOSS:TABLe[:LOW] <FileName>
This command defines the file name of the conversion loss table to be used for the low
(first) range.
Suffix:
<x>
.
irrelevant
Parameters:
<FileName> String containing the path and name of the file.
Example:
MIX:LOSS:TABL 'mix_1_4'
Specifies the conversion loss table mix_1_4.
Manual operation: See "Conversion Loss" on page 422
[SENSe:]MIXer<x>:LOSS[:LOW] <Average>
This command defines the average conversion loss to be used for the entire low (first)
range.
Suffix:
<x>
.
irrelevant
Setting Basic Measurement Parameters