Remote Commands
R&S
®
FSW
1098User Manual 1173.9411.02 ─ 43
Parameters:
<Offset> <numeric value>, specified in Hz, kHz, MHz or GHz, rounded to
the nearest Hz
*RST: 0 Hz
Default unit: HZ
Example: //Define an offset between generator output frequency and ana-
lyzer frequency
SOUR:EXT:FREQ:OFFS 10HZ
Manual operation: See "(Automatic) Source Frequency (Numerator/Denominator/
Offset)" on page 397
SOURce<si>:EXTernal<gen>:POWer[:LEVel]
<Level>
This command sets the output power of the selected generator.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Level> <numeric value>
*RST: -20 dBm
Default unit: DBM
Example: //Define generator output level
SOUR:EXT:POW -30dBm
Manual operation: See "Source Power" on page 396
SOURce<si>:EXTernal<gen>[:STATe] <State>
This command activates or deactivates the connected external generator.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See "Source State" on page 396
Setting Basic Measurement Parameters