Remote Commands
R&S
®
FSW
1110User Manual 1173.9411.02 ─ 43
Suffix:
<p>
.
Power sensor index
Example:
CAL:PMET2:ZERO:AUTO ONCE;*WAI
Starts zeroing the power sensor 2 and delays the execution of
further commands until zeroing is concluded.
Usage: Event
Manual operation: See "Zeroing Power Sensor" on page 376
CALCulate<n>:PMETer<p>:RELative[:MAGNitude] <RefValue>
This command defines the reference value for relative measurements.
Suffix:
<n>
.
Window
<p> Power sensor index
Parameters:
<RefValue> Range: -200 dBm to 200 dBm
*RST: 0
Default unit: DBM
Example:
CALC:PMET2:REL -30
Sets the reference value for relative measurements to -30 dBm
for power sensor 2.
Manual operation: See "Reference Value" on page 377
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
This command sets the current measurement result as the reference level for relative
measurements.
Suffix:
<n>
.
Window
<p> Power sensor index
Example:
CALC:PMET2:REL:AUTO ONCE
Takes the current measurement value as reference value for rel-
ative measurements for power sensor 2.
Usage: Event
Manual operation: See "Setting the Reference Level from the Measurement Meas -
> Ref" on page 377
CALCulate<n>:PMETer<p>:RELative:STATe <State>
This command turns relative power sensor measurements on and off.
Suffix:
<n>
.
Window
Setting Basic Measurement Parameters