Remote Commands
R&S
®
FSW
1114User Manual 1173.9411.02 ─ 43
ON | 1
Switches the function on
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
Manual operation: See "Meas Time/Average" on page 377
[SENSe:]PMETer<p>:ROFFset[:STATe] <State>
This command includes or excludes the reference level offset of the analyzer for power
sensor measurements.
Suffix:
<p>
.
Power sensor index
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
PMET2:ROFF OFF
Takes no offset into account for the measured power.
Manual operation: See "Use Ref Level Offset" on page 378
[SENSe:]PMETer<p>[:STATe] <State>
This command turns a power sensor on and off.
Suffix:
<p>
.
Power sensor index
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
PMET1 ON
Switches the power sensor measurements on.
Manual operation: See "State" on page 376
See "Select" on page 376
[SENSe:]PMETer<p>:UPDate[:STATe] <State>
This command turns continuous update of power sensor measurements on and off.
If on, the results are update even if a single sweep is complete.
Setting Basic Measurement Parameters