Remote Commands
R&S
®
FSW
1131User Manual 1173.9411.02 ─ 43
RMS
RMS value
AVERage
Average
CAVerage
CISPR Average (CISPR filter only)
CRMS
CISPR RMS (CISPR filter only)
*RST: APEak
Example:
DET POS
Sets the detector to "positive peak".
Manual operation: See "Detector" on page 584
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO <State>
This command couples and decouples the detector to the trace mode.
Suffix:
<n>
.
Window
<t> Trace
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
DET:AUTO OFF
The selection of the detector is not coupled to the trace mode.
Manual operation: See "Detector" on page 584
TRACe<n>:COPY <TraceNumber>, <TraceNumber>
This command copies data from one trace to another.
Suffix:
<n>
.
Window
Parameters:
<TraceNumber> TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6
The first parameter is the destination trace, the second parame-
ter is the source.
(Note the 'e' in the parameter is required!)
Example:
TRAC:COPY TRACE1,TRACE2
Copies the data from trace 2 to trace 1.
Manual operation: See "Copy Trace" on page 586