Remote Commands
R&S
®
FSW
1198User Manual 1173.9411.02 ─ 43
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe]......................................... 1199
[SENSe:]DEMod:SQUelch:LEVel................................................................................... 1199
[SENSe:]DEMod:SQUelch[:STATe].................................................................................1199
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous <State>
This command turns continuous demodulation of the signal at the marker position in
the frequency domain on and off.
In the time domain continuous demodulation is always on.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC2:MARK3:FUNC:DEM:CONT ON
Switches on the continuous demodulation.
Manual operation: See "Continuous Demodulation" on page 547
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff <Duration>
This command defines for how long the the signal at the marker position is demodula-
ted.
In the time domain continuous demodulation is always on.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Duration> Range: 10 ms to 1000 s
*RST: Marker demodulation = OFF
Default unit: S
Example:
CALC:MARK:FUNC:DEM:HOLD 3s
Manual operation: See "Marker Stop Time" on page 547
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect <DemodMode>
This command selects the demodulation mode at the marker position.
Suffix:
<n>
.
Window
Analyzing Measurements (Basics)