Remote Commands
R&S
®
FSW
1241User Manual 1173.9411.02 ─ 43
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256 MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see Chapter 5.1.7, "Protecting Data Using the Secure User Mode",
on page 43.
Suffix:
<1|2>
.
irrelevant
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current instrument settings in the file Save.dfl.
MMEM:STOR:STAT:NEXT
Saves the current instrument settings in the file Save_001.dfl
MMEM:STOR:STAT:NEXT
Saves the current instrument settings in the file Save_002.dfl
Manual operation: See "Save File" on page 631
MMEMory:STORe<1|2>:TYPE <arg0>
This command defines whether the data from the entire instrument or only from the
current channel is stored with the subsequent MMEM:STOR... command.
Suffix:
<1|2>
.
irrelevant
Parameters:
<arg0> INSTrument | CHANnel
INSTrument
Stores data from the entire instrument.
CHANnel
Stores data from an individual channel.
*RST: INST
Example:
INST:SEL 'SPECTRUM2'
Selects channel'SPECTRUM2'.
MMEM:STOR:TYPE CHAN
Specifies that channel data is to be stored.
SYSTem:PRESet
This command presets the R&S FSW. It is identical to *RST.
Example:
SYST:PRES
Usage: Event
Managing Settings and Results