Measurements and Results
R&S
®
FSW
169User Manual 1173.9411.02 ─ 43
Number of channels: Tx, Adj.......................................................................................170
Reference Channel..................................................................................................... 171
Noise Cancellation...................................................................................................... 171
Fast ACLR...................................................................................................................172
Selected Trace............................................................................................................ 172
Absolute and Relative Values (ACLR Mode).............................................................. 172
Channel power level and density (Power Unit)........................................................... 172
Power Mode................................................................................................................173
Setting a fixed reference for Channel Power measurements (Set CP Reference)..... 173
Optimized Settings (Adjust Settings)...........................................................................173
Sweep Time................................................................................................................ 173
Standard
The main measurement settings can be stored as a standard file. When such a stan-
dard is loaded, the required channel and general measurement settings are automati-
cally set on the R&S FSW. However, the settings can be changed. Predefined stand-
ards are available for standard measurements, but standard files with user-defined
configurations can also be created.
Note: If the "Multi-Standard Radio" standard is selected, the "ACLR Setup" dialog box
is replaced by the "MSR ACLR Setup" dialog box (see Chapter 7.2.5, "MSR ACLR
Configuration", on page 177).
If any other predefined standard (or "NONE") is selected, the "ACLR Setup" dialog box
is restored (see Chapter 7.2.4, "Channel Power Configuration", on page 168).
Note that changes in the configuration are not stored when the dialog boxes are
exchanged.
Predefined Standards ← Standard
Predefined standards contain the main measurement settings for standard measure-
ments. When such a standard is loaded, the required channel settings are automati-
cally set on the R&S FSW. However, you can change the settings.
The predefined standards contain the following settings:
●
Channel bandwidths
●
Channel spacings
●
Detector
●
Trace Average setting
●
Resolution Bandwidth (RBW)
●
Weighting Filter
For details on the available standards, see Chapter 7.2.9, "Reference: Predefined CP/
ACLR Standards", on page 206.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet on page 842
User Standards ← Standard
Access: "CP / ACLR Config" > "General Settings" tab > "Manage User Standards"
Channel Power and Adjacent-Channel Power (ACLR) Measurement