EasyManuals Logo

R&S FSW Series User Manual

R&S FSW Series
1450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #731 background imageLoading...
Page #731 background image
Network and Remote Operation
R&S
®
FSW
731User Manual 1173.9411.02 ─ 43
Table 13-1: Remote control interfaces and protocols
Interface
Protocols, VISA
*)
address string
Remarks
Local Area
Network
(LAN)
HiSLIP High-Speed LAN Instrument Protocol (IVI-6.1)
TCPIP::host address::hislip0[::INSTR]
VXI-11
TCPIP::host address::inst0[::INSTR]
Library: VISA
socket communication (Raw Ethernet, simple Telnet)
TCPIP::host address[::LAN device name]::<port>::
SOCKET
Library: VISA or socket controller
A LAN connector is located on the
rear panel of the instrument.
The interface is based on TCP/IP and
supports various protocols.
For a description of the protocols refer
to:
VXI-11 Protocol
HiSLIP Protocol
Socket Communication
GPIB (IEC/
IEEE Bus
Interface)
VISA
*)
address string:
GPIB::primary address[::INSTR]
(no secondary address)
A GPIB bus interface according to the
IEC 625.1/IEEE 488.1 standard is
located on the rear panel of the instru-
ment.
For a description of the interface refer
to 13.1.1.2 GPIB Interface (IEC 625/
IEEE 418 Bus Interface).
USB
VISA
*)
address string:
USB::<vendor ID>::<product_ID>::<serial_number>[::INSTR]
USB connectors are located on the
front and rear panel of the instrument.
For a description of the interface refer
to 13.1.1.3 USB Interface.
*
) VISA is a standardized software interface library providing input and output functions to communicate with instruments. A VISA
installation on the controller is a prerequisite for remote control using the indicated interfaces.
(See also Chapter 13.1.3, "VISA Libraries", on page 739).
Within this interface description, the term GPIB is used as a synonym for the IEC/IEEE
bus interface.
13.1.1.1 LAN Interface
To be integrated in a LAN, the instrument is equipped with a LAN interface, consisting
of a connector, a network interface card and protocols. The network card can be oper-
ated with the following interfaces:
10 Mbit/s Ethernet IEEE 802.3
100 Mbit/s Ethernet IEEE 802.3u
1Gbit/s Ethernet IEEE 802.3ab
For remote control via a network, the PC and the instrument must be connected via the
LAN interface to a common network with TCP/IP network protocol. They are connected
using a commercial RJ45 cable (shielded or unshielded twisted pair category 5). The
TCP/IP network protocol and the associated network services are preconfigured on the
instrument. Software for instrument control and the VISA program library must be
installed on the controller.
Remote Control Basics

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals