Remote Commands
R&S
®
FSW
835User Manual 1173.9411.02 ─ 43
INITiate<n>:CONMeas
This command restarts a (single) measurement that has been stopped (using ABORt)
or finished in single sweep mode.
The measurement is restarted at the beginning, not where the previous measurement
was stopped.
As opposed to INITiate<n>[:IMMediate], this command does not reset traces in
maxhold, minhold or average mode. Therefore it can be used to continue measure-
ments using maxhold or averaging functions.
Suffix:
<n>
.
irrelevant
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sweeps) and waits for the
end.
Result: Averaging is performed over 40 sweeps.
Manual operation:
See "Continue Single Sweep" on page 473
INITiate<n>:CONTinuous <State>
This command controls the sweep mode for an individual channel.
Note that in single sweep mode, you can synchronize to the end of the measurement
with *OPC, *OPC? or *WAI. In continuous sweep mode, synchronization to the end of
the measurement is not possible. Thus, it is not recommended that you use continuous
sweep mode in remote control, as results like trace data or markers are only valid after
a single sweep end synchronization.
For details on synchronization see Chapter 13.1.6, "Command Sequence and Syn-
chronization", on page 749.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST: 1
Configuring and Performing Measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals