Remote Commands
R&S
®
FSW
881User Manual 1173.9411.02 ─ 43
Suffix:
<n>
.
1..n
<li> 1..n
<gap> 1 | 2
Gap (CACLR) channel number
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Parameters for setting and query:
<SBGaps> AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
CAL:LIMit:ACPower:GAP2:MAN:UPP:STATe BC,ON
[SENSe:]POWer:ACHannel:BWIDth:GAP<gap>:MANual:LOWer <SBGaps>,
<Bandwidth>
[SENSe:]POWer:ACHannel:BWIDth:GAP<gap>:MANual:LOWer? <SBGaps>
[SENSe:]POWer:ACHannel:BANDwidth:GAP<gap>:MANual:LOWer <SBGaps>,
<Bandwidth>
[SENSe:]POWer:ACHannel:BANDwidth:GAP<gap>:MANual:LOWer? <SBGaps>
Defines the bandwidth of the lower gap channel in the specified gap.
Suffix:
<gap>
.
1 | 2
Gap channel number
Parameters:
<Bandwidth> *RST: 3.84 MHz
Default unit: HZ
Parameters for setting and query:
<SBGaps> AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
POW:ACH:BAND:GAP:MAN:LOW BC,5MHz
Example:
POW:ACH:BWIDth:GAP:MAN:LOW? AB,MIN
Example: See "Example: Configuring and Performing an Asymmetrical
MSR ACLR Measurement" on page 895
Manual operation: See "Gap Channel Bandwidths" on page 192
Configuring and Performing Measurements