Remote Commands
R&S
®
FSW
956User Manual 1173.9411.02 ─ 43
[SENSe:]LIST:RANGe<ri>:DELete
This command removes a range from the sweep list.
Note that
you cannot delete the reference range
a minimum of three ranges is mandatory.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Example:
LIST:RANG2:DEL
[SENSe:]LIST:RANGe<ri>:DETector <Detector>
This command selects the detector for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Detector> APEak
Autopeak
NEGative
minimum peak detector
POSitive
peak detector
SAMPle
sample detector
RMS
RMS detector
AVERage
average detector
*RST:
RMS
Example:
LIST:RANG2:DET AVER
Manual operation: See "Detector" on page 286
[SENSe:]LIST:RANGe<ri>:FILTer:TYPE <FilterType>
This command selects the filter type for a spurious emission measurement range.
The EMI-specific filter types are available if the EMI (R&S FSW-K54) measurement
option is installed, even if EMI measurement is not active. For details see Chap-
ter 7.13.3.1, "Resolution Bandwidth and Filter Types", on page 334.
Configuring and Performing Measurements