Remote Commands
R&S
®
FSW
961User Manual 1173.9411.02 ─ 43
[SENSe:]LIST:RANGe<ri>:SWEep:TIME:AUTO <State>
This command turns automatic selection of the sweep time for a spurious emission
measurement range on and off.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
LIST:RANG2:SWE:TIME:AUTO ON
Manual operation: See "Sweep Time Mode" on page 286
[SENSe:]LIST:RANGe<ri>:TRANsducer <Transducer>
This command selects a transducer factor for a spurious emission measurement
range.
Note that
the transducer must cover at least the span of the range
the x-axis has to be linear
the unit has to be dB
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Transducer> String containing the transducer file name, including the path
information.
Example:
LIST:RANG2:TRAN 'C:\TRANS\MYTRANS'
Manual operation: See "Transducer" on page 287
14.5.8.3 Configuring the List Evaluation
The following commands configure the list evaluation.
Useful commands for spurious emission measurements described elsewhere
MMEMory:STORe<n>:LIST on page 1253
Remote commands exclusive to spurious emission measurements
CALCulate<n>:PSEarch:AUTO....................................................................................... 962
CALCulate<n>:PEAKsearch:AUTO..................................................................................962
CALCulate<n>:ESPectrum:PSEarch:DETails.................................................................... 962
CALCulate<n>:ESPectrum:PEAKsearch:DETails...............................................................962
CALCulate<n>:PSEarch:MARGin.................................................................................... 962