Remote Commands
R&S
®
FSW
979User Manual 1173.9411.02 ─ 43
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd <State>
This command switches on or off the peak-hold function for the active power measure-
ment in zero span in the window specified by the suffix <n>. If activated, the peak for
each sweep is compared to the previously stored peak; the maximum of the two is
stored as the current peak.
The peak-hold function is reset by switching it off and on again.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary[:STATe]
<State>
This command turns time domain power measurements on and off. This measurement
is only available in zero span.
When you turn the measurement on, the R&S FSW activates a marker and positions it
on the peak power level in the marker search range.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN[:STATe] <State>
This command turns the evaluation to determine the mean time domain power on and
off.
The R&S FSW performs the measurement on the trace marker 1 is positioned on.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See "Results" on page 310
Configuring and Performing Measurements