Chapter 5 - Components and Functions (Theory) 5-93
Disruptive Mode
Disruptive Mode provides a way to enable and disable Disruptive mode. When enabled, Disruptive
mode allows a remote user to access the LOGIQ P9/P7 which can disrupt normal operation of the
LOGIQ P9/P7. Disruptive mode can be remotely turned on when the logged-in user has Authorize
Remote Service Access privilege. If the logged-in user does not have remote service privilege, displays
this error message: “You do not have the required permission to perform this operation.”
When initiating Disruptive mode from a remote location, a message displays on the LOGIQ P9/P7
asking whether the user will allow Disruptive mode and notifying them that normal operation might be
disturbed.
•If Yes is selected, then Disruptive mode is turned on.
•If No is selected, then Disruptive mode is not turned on and an error message displays saying that
permission to change the mode was denied.
When Disruptive mode is On, the GE icon in the status bar changes to red and displays this message:
"Due to Service testing restart needed."
When Disruptive mode is changed to Off, the GE icon turns white and displays this message: "Due to
Service testing restart needed." In addition, VCO and SSH stop if running.
The information on Disruptive Mode is available to Class C (Service Advanced) and Class M licenses.
To access Disruptive Mode, select Utility (second page) > Service > Utilities > Disruptive Mode
Utility.
Figure 5-62 Disruptive Mode
For more information:
• Utilities on page 5-73