Data Transfer
Data Transfer provides a way to do the following:
View information about past transfers of (APM) information.
Set up automatic/scheduled transfer of allowed data files from the LOGIQ P9/P7 to the server.
Manually transfer data files allowed data files from the LOGIQ P9/P7 to the server.
The information on Data Transfer is available to all service class licenses.
To access Data Transfer, select Utility (second page) > Service > Utilities > Data Transfer.
Figure 5-63 Data Transfer
This table shows all the elements available on Data Transfer with descriptions.
Table 5-40 Data Transfer
Element DESCRIPTION
Type of Upload
Type of log file. For example, Incremental Logs or Full Logs. Monitoring Logs,
System Logs, and Windows Logs are incrementally transferred when automatic
transfer is enabled. To enable automatic transfer, navigate to System Admin and,
under Service, check Enable Automatic Request for Service.
Upload Permission Whether the permission to upload the log file is allowed or not.
Last Upload Status Whether the last log file upload was successful or not.
Last Upload Attempt Date and time the last log file upload was attempted.
Last Successful Upload Date and time the last log file was successfully uploaded.
Scheduler
When selected, enables the related day selections. For example, All Days,
Monday, and Tuesday.
Save Settings Saves the information.
Send All Manually send the selected log files to the server.

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
ElastographyYes
ConnectivityDICOM, USB, Ethernet, HDMI
Doppler ModesColor Doppler, Power Doppler
Probesconvex, linear, phased array
Imaging ModesM-Mode, Color Doppler, Power Doppler
PortabilityCart-based
ApplicationsVascular, Cardiac, OB/GYN, Musculoskeletal
WeightApprox. 80 kg
Touch Panel10.1-in touchscreen

Related product manuals