GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-201
12.) Unscrew 6 screws and carefully remove the OPIO base.
13.)Remove the IGUS THRUST BUSH.
CAUTION
When reassemble the ROTATION MODULE ASSY, HSH screw should be tightened proper
torque (60kgf-cm)
CAUTION
When reassemble the screws, please use SCREW HSH M5X16 NYLOK(5729013).
CAUTION
When reassemble the OPIO base, HSH screw should be tightened proper touque (60kgf-cm).
CAUTION
When reassemble the screws, please use SCREW HSH M5X16 NYOK (5729013).
CAUTION
DO NOT wipe out grease from part.
IGUS THRUST BUSH