EasyManua.ls Logo

Hioki 8847 - 8.9 Copying settings to other channels (calculation No.) (Copy function)

Hioki 8847
372 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.5 Converting Input Values (Scaling Function)
129
8
Chapter 8 Utility Functions
2
3
4
1
2
4
3
4
Procedure
To open the screen: Press the CHAN key [Each Ch] sheet
1
Enable the Scaling function.
Select
Move the cursor to the [Disp] item.
2
Select the scaling conversion method.
Select
Move the cursor to the [Method] item.
3
Specify the physical units.
Move the cursor to the [Unit], and enter the physi-
cal unit name. (Up to 7 characters)
See: "Entering Text" ( p.121)
4
Enter the numerical values for conversion.
When you have selected [Ratio] (set conver-
sion ratio and offset)
Move the cursor to the [Ratio] and [Offset] items.
Enter numerical values in each field.
When you have selected [2-Point] (set input val-
ues for two points and the values after conver-
sion)
Move the cursor to the [Input P1], [Scale P1],
[Input P2], and [Scale P2].
Enter numerical values in each field.
Off No scaling.
NUM
Displayed as a decimal and includes a unit (m,
k, etc.).
SCI
Displayed as an exponent (the power of ten)
Ratio Specify by conversion ratio.
Point
Specify by two points.
-9.9999E+9 to 9.9999E+9
-9.9999E+9 to 9.9999E+9
Input example:
Decimal1.2345 mV
Exponent1.2345E-03V
Display selecting [Ratio] for [Method] item
Display selecting [Point] for [Method] item
When saving text or results of numerical calculation, some characters and
symbols used for display on the instrument will be converted as follows. (8847
display saved string)
2
^2,
3
^3 , μ → ~u, Ω→ ~o, ε→ ~e, °~c ,
±~+, με (display only)uE, °C (display only)C
To enter the current input value as is for P1 or P2
Select [Monitor Val].

Table of Contents