8.5 Converting Input Values (Scaling Function)
129
8
Chapter 8 Utility Functions
Procedure
To open the screen: Press the CHAN key →[Each Ch] sheet
1
Enable the Scaling function.
Move the cursor to the [Disp] item.
2
Select the scaling conversion method.
Move the cursor to the [Method] item.
3
Specify the physical units.
Move the cursor to the [Unit], and enter the physi-
cal unit name. (Up to 7 characters)
See: "Entering Text" ( p.121)
4
Enter the numerical values for conversion.
When you have selected [Ratio] (set conver-
sion ratio and offset)
Move the cursor to the [Ratio] and [Offset] items.
Enter numerical values in each field.
When you have selected [2-Point] (set input val-
ues for two points and the values after conver-
sion)
Move the cursor to the [Input P1], [Scale P1],
[Input P2], and [Scale P2].
Enter numerical values in each field.
Off No scaling.
NUM
Displayed as a decimal and includes a unit (m,
k, etc.).
SCI
Displayed as an exponent (the power of ten)
Ratio Specify by conversion ratio.
Point
Specify by two points.
Input example:
Decimal1.2345 mV
Exponent1.2345E-03V
Display selecting [Ratio] for [Method] item
Display selecting [Point] for [Method] item
• When saving text or results of numerical calculation, some characters and
symbols used for display on the instrument will be converted as follows. (8847
display → saved string)
•
2
→ ^2,
3
→ ^3 , μ → ~u, Ω→ ~o, ε→ ~e, °→ ~c ,
±→ ~+, με (display only) →uE, °C (display only)→ C
To enter the current input value as is for P1 or P2
Select [Monitor Val].