13.3 Setting FFT Analysis Conditions
224
The display range of the vertical (y) axis can be set to automatically suit analysis results, and can be freely
expanded and compressed.
13.3.10Setting the Display Range of the Vertical Axis
(Scaling)
1
Select automatic or manual scaling of the y-axis display.
Move the cursor to the [Scale] column of the Analysis No. to set.
2
When [Manu] is selected
Set the upper and lower limits to display.
Move the cursor to the
[Lower] or [Upper] item.
Set the upper and lower limits to display the analysis results.
Setting range: -9.9999E+29 to +9.9999E+29
(with exponent from E-29 to E+29)
See: "8.1.3 Alphanumeric Input" ( p.121)
Auto Scaling of the vertical (y) axis is automatically set according to
analysis results. (default setting)
Manu-
al
Scaling of the vertical (y) axis can be set as desired, to suit the pur-
pose of the measurement.
This is useful for magnifying or reducing the displayed amplitude,
and for shifting the displayed waveform up or down.
Procedure
To open the screen: Press the STATUS key → [Status] sheet
How do I copy settings to other calculation No.?
See: "8.9 Copying settings to other channels (calculation No.) (Copy function)" (
p.139)