EasyManua.ls Logo

Hioki 8847 - 3.4.3 Recording Length (number of divisions)

Hioki 8847
372 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.7 Adjusting the Zero Position (Zero-Adjust)
36
This procedure compensates for input module differences and sets the reference potential of the
instrument to 0 V.
The compensation applies to the selected range.
Perform zero-adjust in the following cases.
When an input module was changed.
When power was turned off and on again.
When settings were initialized (system reset).
When DC/RMS is switched at the 8971 Current Unit and the 8972 DC/RMS
Unit
When the ambient temperature has changed significantly.
Zero-position drift
*
may occur.
* Drift: This refers to spurious output caused by a shift in the operating point of an opera-
tional amprifier. Drift can occur due to changes in temperature and due to compo-
nent aging over a period of use.
2.7 Adjusting the Zero Position (Zero-Adjust)
To obtain precise measurements, provide about 30 minutes warm-up after turning power on to allow
the internal temperature of the input modules to stabilize.
Note that zero-adjust cannot be performed during a measurement.
During zero-adjust, the operation keys of the unit are inactive. (The procedure may take several sec-
onds.)
Before starting zero-adjust
Procedure
To open the screen: Press the CHAN key [Unit List] sheet
1
Move the cursor to the [Zero-Adjust] item.
2
Select [Exec Zero-Adjust].
The zero-adjust procedure is carried out.
Zero-adjust has no effect on the 8969 Strain Unit. (Perform zero-adjust using
Auto Balance. ( p.144))

Table of Contents

Other manuals for Hioki 8847