10.1 Numerical Calculation Workflow
174
There are two different ways of performing calculation.
• Calculating While Measuring: Settings for numerical calculation must be made before the measurement.
• Applying Calculations to Existing Data: Calculation is possible for waveform data after measurement is
completed, and for data saved on media.
10.1 Numerical Calculation Workflow
Calculating While Measuring
Make Calculation Settings Make calculation settings on the [Num Calc] sheet. ( p.176)
When judging based on calculation results:
To automatically print or save calculation results: make printing
and saving settings before measuring.
• Saving calculation results automatically
• Printing calculation results automatically
(
p.180)
(
p.183)
(
p.184)
Start Measurement
Acquire Data The instrument acquires data when the trigger criteria are
met.
(If triggering is not enabled, the instrument acquires data when
you press the START key.)
Calculate "Measurement" appears on the screen's status bar.
Calculations are performed sequentially from No. 1 to No. 16.
(To interrupt calculations, press the STOP key to abort)
Display Calculation Results Results are displayed on the Waveform screen. ( p.179)
(Calculation Judgments) (only if judgments are enabled)
(Print or Save) (if auto printing or auto saving are enabled)
Calculation results are automatically printed or saved.
Stop Measurement